Ivium Technologies  |  SKU: pocketSTAT2

Potentiostat pocketSTAT2

€0,00


Delivery and Shipping to EU

We will add in the quotation, the shipping, insurance, customs clearance costs.

Potentiostat pocketSTAT2

Ivium Technologies

The pocketSTAT2 is a portable potentiostat/galvanostat/ZRA with integrated FRA/EIS, and has the size and weight of a smart phone. It is controlled via a USB connection from any netbook, laptop or PC that is Windows operated.

Key specifications 
– Size: 16 x 6.7 x 1.9cm
– Weight: 300g
– Scan range: ±10V/±30mA
– 4 Electrode connection: WE/CE/RE/S and GND lead
– Max. acquisition rate: 300,000 pnts/s

Rugged design
The pocketSTAT2 has a housing made of strong, yet lightweight aluminium.

Hardware

    Automatic current range selection: 100pA to 10mA/1mV to 10V
    Automatic variable noise filter and potentiostat/galvanostat stability settings
    Real time data up to 3,000 pnts/s. Acquisition up to 300,000 pnts/s is stored in instrument memory
    Applicable scanrates: 1µV/s to 10,000V/s
    Minimum interval time: 3.33µs (0.125µs resolution)
    Result optimalisation by automated tuned filters, gain amplifiers and DC-subtraction

Software
The pocketSTAT2 is controlled via Ivium’s own IviumSoft. This versatile, yet intuitive software package allows instrument control, data management and analysis, etc. IviumSoft can also interface with and be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of IviumSoft is included with each instrument as standard.

System Performance
Current compliance ±30mA
Maximum compliance voltage ±10V
Maximum applied voltage ±10V
Electrode connections 4; WE, CE, RE, S (and GND) with 4mm banana plugs
Potentiostat bandwidth >500kHz
Stability settings High Speed, Standard and High Stability
Programmable response filter 1MHz, 100kHz, 10kHz, 1kHz, 10Hz
Signal acquisition Dual channel 18bit ADC, 300,000 samples/s stored in instrument memory (max. 1M datapoints)
Potentiostat
Applied potential range ±10V, 0.08mV res.
Applied potential accuracy 0.2%, or 2mV
Current ranges ±100pA to ±10mA in 9 decades
Measured current resolution 0.003% of current range, minimum 3fA
Measured current accuracy 20pA + 0.025% of FSR
Galvanostat
Applied current resolution 0.008% of applied current range
Applied current accuracy 0.2%
Galvanostatic current ranges ±10nA to ±10mA
Potential ranges ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
Measured potential resolution 0.0008% of potential range, minimum 7nV
Measured potential accuracy 0.2% or 2mV
Impedance Analyser
Frequency range 10µHz to 1MHz
Amplitude 0.15mV to 2.0V, or 0.03% to 100% of current range
DC offset 16bit DC offset subtraction and 2 DC-decoupling filters
Electrometer
Input impedance >1000Gohm //<10pF
Input bias current <20pA
Bandwidth >5MHz
Special functions
Ohmic drop compensation 2V/current range
Peripheral connections
Analog in/out ±10V input only, 18bit resolution, bandwidth 40kHz (*not available with cell cable delivered as standard)
Environment
Power requirements Via USB
Interfacing USB
Size (w x d x h) 16 x 6.7 x 1.9cm
Weight 300g
PC requirements
Operating System Windows 8/10/11, with free USB port

 

Datasheet PDF pocketSTAT2