BaSyTec  |  SKU: CTS Standard

BaSyTec CTS Battery Test System

€0,00


Delivery and Shipping to EU

We will add in the quotation, the shipping, insurance, customs clearance costs.

BaSyTec CTS Battery Test System

BaSyTec

The BaSyTec CTS is the reference when testing small cells from material probes in 3 electrode arrangement up to mid size cells up to 20A.
Of course also our CTS is controlled by our powerful and comfortable BaSyTest Software.
The 4 automatic dynamic switching current ranges (switching without interruption) (plus paralleling) result in a relative accuracy of at least 0.33% within the whole range from
60μA to 20A.

Main features:

  • 1-6V or +/-6V
  • up to 5A (20A)
  • up to 200nA current accuracy
  • optional reference electrode input

PDF Datasheet

Technical data

  CTS Standard CTS LAB CTS LAB XL
Channels per unit (max.) 32 20 32
Voltage range 0-6V +/-6V +/-6V
4-wire measurement Yes Yes Yes
Min. discharge voltage 200mV or 0.2Ω*current (whichever is greater) 200mV or 0.2Ω*current (whichever is greater) 200mV or 0.2Ω*current (whichever is greater)
Current ranges (charge + discharge) 5A/300mA/15mA/1mA 3A/250mA/15mA/1mA 5A/300mA/15mA/1mA
Current accuracy 1mA/50µA/2.5µA/0.2µA 1mA/50µA/2.5µA/0.2µA 1mA/50µA/2.5µA/0.2µA
Voltage accuracy 1mV 1mV 1mV
Resolution 16Bit / 1us 16Bit / 1us 16Bit / 1us
Parallel operation Yes, up to 4 channels Yes, up to 4 channels Yes, up to 4 channels
Additional inputs per channel NTC temperature (for Epcos B57861S0502) NTC temperature (for Epcos B57861S0502) NTC temperature (for Epcos B57861S0502)
Optional reference voltage +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate
Digital-IO Optional, TTL-like, 3 per channel, not isolated Optional, TTL-like, 3 per channel, not isolated Optional, TTL-like, 3 per channel, not isolated
Sampling rate Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system
Current rise time < 50µs < 50µs < 50µs
3-Electrode configurations Limited Yes Yes
Control mode Current, voltage, power, resistance and combinations Current, voltage, power, resistance and combinations Current, voltage, power, resistance and combinations
Other methods Cyclic voltammetry, GITT, … Cyclic voltammetry, GITT, … Cyclic voltammetry, GITT, …
Software BaSyTest BaSyTest BaSyTest
Interface Ethernet Ethernet Ethernet
Battery connectors D-SUB 9-pin D-SUB 9-pin D-SUB 9-pin
Holders for cylindric cells Available Available Available
Size per unit (max.) 19” system 6HE units, 450280550mm 19” system 6HE units, 450280550mm 19” system 6HE units, 450280550mm
Weight per unit (max.) 28kg (32 channels) 28kg (32 channels) 28kg (32 channels)
Noise < 55dB(A) < 55dB(A) < 55dB(A)
Input power / voltage range < 1700W / 110-264VAC / 47-63Hz < 1700W / 110-264VAC / 47-63Hz < 1700W / 110-264VAC / 47-63Hz
Max. input current per unit < 16A @ 110V, < 8A @ 220V < 16A @ 110V, < 8A @ 220V < 16A @ 110V, < 8A @ 220V