BaSyTec CTS Battery Test System
Couldn't load pickup availability
Delivery and Shipping to EU
Delivery and Shipping to EU
We will add in the quotation, the shipping, insurance, customs clearance costs.
BaSyTec CTS Battery Test System
BaSyTec
The BaSyTec CTS is the reference when testing small cells from material probes in 3 electrode arrangement up to mid size cells up to 20A.
Of course also our CTS is controlled by our powerful and comfortable BaSyTest Software.
The 4 automatic dynamic switching current ranges (switching without interruption) (plus paralleling) result in a relative accuracy of at least 0.33% within the whole range from
60μA to 20A.
Main features:
- 1-6V or +/-6V
- up to 5A (20A)
- up to 200nA current accuracy
- optional reference electrode input
Technical data
CTS Standard | CTS LAB | CTS LAB XL | |
Channels per unit (max.) | 32 | 20 | 32 |
Voltage range | 0-6V | +/-6V | +/-6V |
4-wire measurement | Yes | Yes | Yes |
Min. discharge voltage | 200mV or 0.2Ω*current (whichever is greater) | 200mV or 0.2Ω*current (whichever is greater) | 200mV or 0.2Ω*current (whichever is greater) |
Current ranges (charge + discharge) | 5A/300mA/15mA/1mA | 3A/250mA/15mA/1mA | 5A/300mA/15mA/1mA |
Current accuracy | 1mA/50µA/2.5µA/0.2µA | 1mA/50µA/2.5µA/0.2µA | 1mA/50µA/2.5µA/0.2µA |
Voltage accuracy | 1mV | 1mV | 1mV |
Resolution | 16Bit / 1us | 16Bit / 1us | 16Bit / 1us |
Parallel operation | Yes, up to 4 channels | Yes, up to 4 channels | Yes, up to 4 channels |
Additional inputs per channel | NTC temperature (for Epcos B57861S0502) | NTC temperature (for Epcos B57861S0502) | NTC temperature (for Epcos B57861S0502) |
Optional reference voltage | +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate | +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate | +/-6V, 1mV accuracy, leakage < 50nA, 200ms samplerate |
Digital-IO | Optional, TTL-like, 3 per channel, not isolated | Optional, TTL-like, 3 per channel, not isolated | Optional, TTL-like, 3 per channel, not isolated |
Sampling rate | Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system | Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system | Peak 0.5ms, Min. 10ms but not more than 1kHz continually per system |
Current rise time | < 50µs | < 50µs | < 50µs |
3-Electrode configurations | Limited | Yes | Yes |
Control mode | Current, voltage, power, resistance and combinations | Current, voltage, power, resistance and combinations | Current, voltage, power, resistance and combinations |
Other methods | Cyclic voltammetry, GITT, … | Cyclic voltammetry, GITT, … | Cyclic voltammetry, GITT, … |
Software | BaSyTest | BaSyTest | BaSyTest |
Interface | Ethernet | Ethernet | Ethernet |
Battery connectors | D-SUB 9-pin | D-SUB 9-pin | D-SUB 9-pin |
Holders for cylindric cells | Available | Available | Available |
Size per unit (max.) | 19” system 6HE units, 450280550mm | 19” system 6HE units, 450280550mm | 19” system 6HE units, 450280550mm |
Weight per unit (max.) | 28kg (32 channels) | 28kg (32 channels) | 28kg (32 channels) |
Noise | < 55dB(A) | < 55dB(A) | < 55dB(A) |
Input power / voltage range | < 1700W / 110-264VAC / 47-63Hz | < 1700W / 110-264VAC / 47-63Hz | < 1700W / 110-264VAC / 47-63Hz |
Max. input current per unit | < 16A @ 110V, < 8A @ 220V | < 16A @ 110V, < 8A @ 220V | < 16A @ 110V, < 8A @ 220V |