Working Voltage |
- AC 220V 50/60 Hz Single Phase
|
Max Power Consumption |
|
Glove Box Chamber

|
- Material: Stainless Steel 304
- Dimensions: 4800 mm (L) x 800 mm (W) x 930 mm (H)
- 12 mm thick double layer laminated safety glass with corrosion protection film for operating in a safe and easy manner
- Ample space for the cell fabrication line
|
Chamber Condition |
- Water concentration: < 1 ppm (20°C, 1 atm)
- Oxygen concentration: < 1 ppm (20°C, 1 atm)
|
Leak Rate |
- <0.05 Vol% / hr (Depending on oxygen concentration inside the glove box)
|
Airlock Chamber |
- Two Antechambers and two small chambers on both sides of the glove box
- Antechamber Dimension: 385 mm (Dia.) x 588 mm (L)
- Small Chamber Dimension: 150 mm (Dia.) x 300 mm (L)
- Maximum vacuum level: 100 Pa
|
Working Gas |
- Operating gas: Inert gas such as N2, Ar, He
- Regeneration gas: Mixture of H2(5%) and Operating gas
- For purification system with only water adsorbing function, Regenerating gas would be the same as Operating gas
|
Filter System |
- Two Purification Cylinders included due to the large glove box
- Organic Solvent Filter System filled with replaceable activated carbon
- Built in HEPA filter can eliminate particle size > 0.3?m
|
Control System |
- Siemens PLC control system with 6" touchscreen display
- Water concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm
- Oxygen concentration: 0 - 1000 ppm with an accuracy of 0.1 ppm
- Pressure: -2500 to 2500 Pa with an accuracy of 1Pa
- Purification limitation
- Water concentration: <1 ppm (20°C, 1 atm)
- Oxygen concentration: <1 ppm (20°C, 1 atm)
|
Pump

|
- 240 L/m Heavy Duty Double Stage Rotary Vane Pump for the gas exhaust
- Vacuum rate: 7.8 CFM
- Lowest pressure: 04 Pa
|
Blower |
|
Purification Pipeline |
- 304 Stainless Steel, KF40 Port
|
Gloves |
- A pair of NORTH Butyl glove is included to provide better protection against toxic materials for one chamber
- Four pairs in total
|
Lighting |
- Built-in fluorescent lamp
|
Slidable Shelf
|
- Four slidable shelves are built on top of inside glove chamber with dimensions of 300mm x 250mm
- 300 mm away from the roof
|
Optional |
- Universal terminal panel is a convenient part with varies feedthrough built-in for VGB series glove boxes to transfer gas/vacuum pipelines, circuit cables, etc.
- Dimensions: 420x280mm
- Adapter: 10 ?6 ports, 4 ?8 ports, 4 ?12 ports
- Heatable Antechamber: 200C max., DN385×588mm (L), extra cost $7699

|
Product Dimensions |
- 6500mm(L) x 830mm(W) x 1830mm(H) (256" x 33" x 72")
|
Shipping Dimension & Weight |
|
Operation Instructions |
|
Warranty |
- One-year limited warranty with lifetime support
- Rusting and damage due to the improper storage condition or maintenance would not be covered by warranty
- Gloves are consumable parts and are not covered by warranty
|
Application Notes & Warnings |
- The interconnections between the glove box chamber and the gas purification system must be unimpeded during the purification cycles.
- The use of corrosive gasses is prohibited as they will damage the water and oxygen sensors.
- Regularly perform regeneration to maintain the optimum purification efficiency.
- The O2 removing rate is largely dependent on the type of purging gas used. For faster chamber purging process, Nitrogen is preferred over Argon due to its lighter mass.
- Corrosive liquid (such as LiPF6 electrolyte) must be remain sealed in a container inside the glove box. Otherwise, liquid vapor may condense and corrode the steel chamber and purification pipeline
|